ANESTHESIA MACHINE INTEGRATED FOR PEDIATRIC AND ADULT
Standard Features:
- With 10.4 TFT color screen
- Mounted on 4 castors. 2 front castors with brakes.
- With 1 safety lock and 3 big drawers.
- Handle for easy maneuvering.
- Aluminium alloy rail frame. With durable and non-corrosive stainless steel platform.
- Equipped with high precise Anesthesia Vaporizer (with temperature, flow and pressure compensation function). The agent output is continuous, stable & precise. The agent type: Isoflurane / Enflurane / Halothane / Sevoflurane (Select any two)
- With Volume & pressure ventilation modes.
- Classical CO2 circle absorber with 2 canisters for low resistance and minimum gas leakage.
- Fresh gas outlet is in front of frame for the user to easily connect Jackson-Rees circuit.
- With upgraded with functions, eg, CO2 Monitor, Multi-parameter monitor, Anesthesia depth monitor, etc.
- Working interfaces for adult and pediatric are separated; the doctor can choose interface according to patient type.
Optional : O2, N2O Yoke device. When main gas supply is fails, The backup gas supply starts.
Features : | |
---|---|
Application : | Adult, Pediatric |
Driven mode : | Pneumatically driven electronically controlled |
Mode : | Semi-open, Semi-closed or closed system |
Anesthetic vaporizer : | Two high precise vaporizers, Isoflurane, Enflurane, Halothane, Sevoflurane. |
Power supply : | 220V±10%, 50Hz ±1Hz |
Internal battery : | DC12V (over 4 working hours) |
Fresh gas outlet is in front of frame | |
Anesthesia Ventilator : | |
Respiration mode : | IPPV, VCV, PCV, SIMV, SIGH, MANUAL, STANDBY, DEMO |
Waveform : | P-t, F-t, V-t, F-v, V-p |
Tidal Volume : | 20~1600ml |
Respiratory frequency : | 1 ~ 65 BPM |
I: E ratio: | 4: 1~1: 4 |
Inspiration pause : | 0~55%Ti |
Inspiration Pressure : | 1~60cmH2O |
Trigger Pressure : | -20~20cmH2O |
Monitoring parameters : | O2 concentration, MV, Respiratory frequency, Tidal volume, Ppeak, PEEP, I: E ratio, Pmean, etc. |
Alarm parameters : | High pressure alarm, low pressure alarm, O2 concentration alarm, MV alarm, O2 source failure alarm, AC power failure alarm, etc. |
Pressure alarm setting : | High pressure alarm set 1~60cmH2O Low pressure alarm set 0~20cmH2O |
O2 concentration alarm setting: | high 21~99% low 21~50% |
MV alarm setting : | MV high alarm set 10~99L/min MV low alarm set 0~40L/min |
Main Unit: | |
Circle absorber : | Compact circle absorber |
Flow meter : | Five~tube flow meter O2 tube: 0L/min~1L/min~10L/min. N2O tube: 0L/min~1L/min~10L/min. Air tube: 0L/min~10L/min O2 and N2O Linkage Device |
Keep oxygen concentration over 25% | |
Oxygen Flush : | 25L/min ~ 75L/min |
Gas source pressure : | O2 : 0.25Mpa ~ 0.65Mpa N2O : 0.25Mpa ~ 0.65Mpa Air: 0.25Mpa ~ 0.65Mpa |
Gauge : | 4 units, 2 x O2, N2 O, Air. |