BLOOD BAG SYSTEMS
NOTE: PVC material - Roll Welding Film
Blow extrude also available, Add "B" at the end of item code
Optional Accessories:
- Needle Protector
- Sample Pouch + Collection Tube
|
|
Size |
Solution |
Single bag |
|
250ml |
CPDA1,39ML |
Needle -16G |
|
350ml |
CPDA1,49ML |
|
|
450ml |
CPDA1,63ML |
|
|
500ml |
CPDA1,70ML |
|
|
Size |
Solution |
Double bag |
|
250ml |
CPDA1,39ML |
Needle -16G |
|
350ml |
CPDA1,49ML |
|
|
450ml |
CPDA1,63ML |
|
|
500ml |
CPDA1,70ML |
|
|
Size |
Solution |
Triple bag |
|
250ml |
CPDA1,39ML |
Needle -16G |
|
350ml |
CPDA1,49ML |
|
|
450ml |
CPDA1,63ML |
|
|
500ml |
CPDA1,70ML |
|
|
Size |
Solution |
Triple bag |
|
250ml |
CPDA1,39ML SAGM-56ML |
Needle -16G |
|
350ml |
CPDA1,49ML SAGM-78ML |
|
|
450ml |
CPDA1,63ML SAGM-100ML |
|
|
500ml |
CPDA1,70ML SAGM-111ML |
|
|
Size |
Solution |
Quadruple bag |
|
250ml |
CPDA1,39ML |
Needle -16G |
|
350ml |
CPDA1,49ML |
|
|
450ml |
CPDA1,63ML |
|
|
Size |
Solution |
Quadruple bag |
|
250ml |
CPDA1,39ML SAGM-56ML |
Needle -16G |
|
350ml |
CPDA1,49ML SAGM-78ML |
|
|
450ml |
CPDA1,63ML SAGM-100ML |
Blood Bag Product Introduction
- High quality blood bags are designed for most secure blood collection, separation, preservation and transfusion.
Standard Features:
- Flexible pre-sterilized,pyrogen, non toxic, non-hemolytic, biocompatible material.
- No risk of contamination and air embolism (closed system).
- Slit on both sides of the bags can accommodate 5-10ml volume test tubes.
- The capacity of the bag is enough to prevent any rupture of the bag from the seam when it is filled to the requisite volume of blood.
- High quality, peel-resistant label provides for convenient writing.
- Ultra thin walled 16G venipuncture needle is optimal for reducing donor pain, optional needle protection system can keep the operator always safe.
- CPDA-1 or CPD/SAG-M anticoagulant preservation for 35 days or 42days whole blood storage or red cell storage respectively.
- Easy to break off valve allows plasma transfer to both directions freely.
- Round bag shape for increasing yield of blood components during separation.
- Sterilized by high pressure, high temperature steam.
- Optional Components:
- Attached blood sampling device.
- Needle protector.
- "Y" injection port.
- In-line leukocyte reduction filter.
- Platelet storage bag (platelet storage for 3 to 7 days).
- Type of outlet ports (twist off or tear off).
- Volume of primary bag and satellite bags.
- Customize Design
Anticoagulant Solution Choices
Solution |
Purpose |
Whole blood/red cell storage |
Additive concentration (per 100ml) |
Performance |
|
CPD |
Anticoagulation and storage of blood |
21 days |
Citric acid (anhydrous) |
0.299g |
Prevent coagulation of blood |
Sodium citrate (dihydrate) |
2.63g |
Adjust pH |
Monobasic sodium phosphate (monohydrate) |
0.222g |
Nutrition source of red cell |
Dextrose (monohydrate) |
2.55g |
Supports ATP production |
Water for injection |
q.s100ml |
|
CPDA 1 |
Anticoagulation and storage of blood |
35 days |
Citric acid (anhydrous) |
0.299g |
Prevent coagulation of blood |
Sodium citrate (dihydrate) |
2.63g |
Adjust pH |
Monobasic sodium phosphate (monohydrate) |
0.222g |
Nutrition source of red cell |
Dextrose (monohydrate) |
3.19g |
Supports ATP production |
Adenine |
0.0275g |
Supports to maintain ATP level |
Water for injection |
q.s100ml |
In red cell |
SAGM |
Red cell preservation |
42 days |
Dextrose (monohydrate) |
0.900g |
Nutrition source of red cell |
Sodium chloride |
0.877g |
Adjust osmotic pressure |
Adenine |
0.0169g |
Supports to maintain ATP level |
D-mannitol |
0.525g |
In red cell |
Water for injection |
q.s.100ml |
Supports integrity of red cell membrane (to avoid haemolysis) |