DIGITAL SINGLE CHANNEL ECG
Specifications/Details :
- Large LCD display waveform and data
- High-resolution thermal printing array system
- Frequency response is as high as 150Hz
- Capable of printing continuously on channel trace and annotations including lead mark and parameter such as sensitivity, paper speed and filter operation status
- High resolution digital filter inhibits baseline drift without affecting ECG waveform
- Concept of floating input circuit meets IEC safety standard
- Rechargeable battery supporting more than 50 patient examinations
Specification:
- Input Circuit Floating; Protection against Defibrillator effect
- Lead: Standard 12 Leads, Lead change automatically
- Patient Current Leakage: ≤10μA
- Calibrating Voltage: 1mV±3%
- Frequency Response: 0.05Hz~150Hz (IEC) CMRR>100dB
- Paper Speed: 25, 50mm/s±3%
- Recording Paper: 50mm,20m, high-speed roll
- Input Circuit Current: ≤50mA
- Sensitivity: 1/2, 1, 2 (cm/mV)
- Skin Voltage Tolerance: ±300mV
- Dimensions: 310mm×230mm×70mm
- Input Impedance: ≥50ΜΩ
- A/D Conversion: 12 bit
- Time Constant: >3.2seconds
- EMG Filter: 35Hz (-3dB) / 25Hz (-3dB
- Safety Standard: IEC class I, type CF
- Noise Level : <15μVp-p
- Weight 2.75kg
Standard Accessories:
- Patient Cable,Thermal Printing Paper, Limb Electrodes, Chest Electrodes, Power Supply Cable, Grounding Wire